Skip to main content

Scissors, Tonsil, Metzenbaum, Curved, 18 cm - 1002023

Product ID: ITM-1002023
generic
GENERAL DESCRIPTION
Scissors, Metzenbaum, 180 mm, curved, blunt / blunt.

INTENDED USE
Surgical scissors used for cutting of dissecting delicate tissue.

PRODUCT DESCRIPTION / TECHNICAL SPECIFICATIONS Static function
See Martin Catalogue no. 24 Ref: 11 255 14.
Instrument that cuts by shearing: surgical scissors.
Thin, curved, with blunt end blades.
Length: 18 cm.
Material: Martensitic steel (quenched, magnetic steel).
Martensitic steel composition: 0.15-0.40 % carbon; 13-14 % chromium.
Reusable.

Packaging and labelling:
Unit presentation: individually, in protective packaging.
The following should appear on the packaging: - Designation of the instrument.
- Name and address and supplier (manufacturer)
Symbols used according ISO 15223 CE Mark

Instructions for use: To retract skin, fatty tissue, muscles of viscera during superficial surgery.
This item must be cleaned, disinfected after each use, and sterilised in a steam steriliser.
Used to grip soft tissue

Regulatory & Conformity Requirements:
CE mark Comply to EU Council Directive 93/42/EEC (MDD) of EU Regulations 2017/745 (MDR) with valid EC Certificate and Declaration of Conformity for the CE mark is applicable.


US FDA 510(K) clearance

CE certificate available for Class IIa and above devices
Manufacturer’s Quality Management System is certified to ISO 13485 : 2016

Classification: Classified under EU MDR 2017/745 as Class I device

Safety & product Standards:
ISO 7153-1:2016: Surgical instruments - Materials - Part 1: Metals. ISO 7151:1988: Surgical
instruments
- Non-cutting, articulated instruments - General requirements and test methods.
ISO 13402:1995: Surgical and dental hand instruments - Determination of resistance against
autoclaving, corrosion and thermal exposure. ISO 17664:2017: Processing of health care products -
Information to be provided by the medical device manufacturer for the processing of medical devices.
ISO 14971:2007 Medical devices - Application of risk management to medical devices.

Environmental conditions: Storage and transport conditions: 1°C to 30°C.
Humidity: 30% RH to 70% RH.

Pack SizeAvg. Price
Pack of 1$2.44