GENERAL DESCRIPTION:
Neonatal mobile resuscitation table, including radiant warmer, overhead light and APGAR timer and X-ray slot including accessories.
INTENDED USE
Newborn resuscitation and support of transition of infants at birth. Additionally, used to provide thermal support for newborns in the delivery suite and for critically ill infants who require constant nursing intervention.
TECHNICAL SPECIFICATIONS
The unit is supported by a sturdy and stable wheeled base.
Bases is provided with a minimum four (4) anti-static swivel castors and at least two (2) of the castors are provided with brakes.
The base is equipped with a side handle/rail for easy positioning.
Integrated support for 2 x 10 L oxygen cylinders.
Side rail and IV pole for mounting accessories and equipment.
Provided with a minimum of one (1) lower drawer for storage.
An X-ray cassette compartment is integrated under the infant bed.
Fixed-height hood above the table integrates heating element and overhead light.
The vertical column integrates controls and a display which visualises working parameters and alarms.
Equipped with a Treatment /CPR timer which allows pre-sets between 1 to 59 minutes in increments of 1 min.
Equipped with an APGAR timer which allows for up and down counting, and provides warning signals after 1, 5 and 10 min.
Power requirements: 100 - 240 Volts - 50/60 Hz (not necessarily in a single unit).
TABLE FEATURES:
Surface 58 W x 71 L cm (minimum guaranteed dimensions), minimum surface area 4,118 cm².
The side panels made of transparent acrylic resin are collapsible/foldable for complete accessibility.
Is equipped with a tiltable cradle, up to at least 10°.
Mattress covering the entire table surface, thickness no less than 5 cm.
Mattress cover flame-retardant, waterproof, and able to withstand hospital-grade disinfectant products.
Warm-up time < 30 min from 25°C.
OVERHEAD HEATER FEATURES:
Overhead radiant heater at a fixed height of 1.80m.
Radiant heating is achieved by quartz/ceramic elements, placed in a parabolic reflector.
Radiator is equipped with a safety grid.
The unit support three (3) modes: Pre-heating, manual and servo mode.
The unit allows for instant switching been the different operating modes.
The minimum skin temperature setting range is 32°C to 37°C.
The temperature control accuracy is: ± 0.5°C and the setting adjustment resolution is 0.1°C.
In manual and pre-heating modes heating settings allow for a range of 0 – 100%.
LED spotlight(s) for examination are integrated in the overhead radiant heater, the angle of the overhead light is adjustable.
Neonatal mobile resuscitation table, including radiant warmer, overhead light and APGAR timer and X-ray slot including accessories.
INTENDED USE
Newborn resuscitation and support of transition of infants at birth. Additionally, used to provide thermal support for newborns in the delivery suite and for critically ill infants who require constant nursing intervention.
TECHNICAL SPECIFICATIONS
The unit is supported by a sturdy and stable wheeled base.
Bases is provided with a minimum four (4) anti-static swivel castors and at least two (2) of the castors are provided with brakes.
The base is equipped with a side handle/rail for easy positioning.
Integrated support for 2 x 10 L oxygen cylinders.
Side rail and IV pole for mounting accessories and equipment.
Provided with a minimum of one (1) lower drawer for storage.
An X-ray cassette compartment is integrated under the infant bed.
Fixed-height hood above the table integrates heating element and overhead light.
The vertical column integrates controls and a display which visualises working parameters and alarms.
Equipped with a Treatment /CPR timer which allows pre-sets between 1 to 59 minutes in increments of 1 min.
Equipped with an APGAR timer which allows for up and down counting, and provides warning signals after 1, 5 and 10 min.
Power requirements: 100 - 240 Volts - 50/60 Hz (not necessarily in a single unit).
TABLE FEATURES:
Surface 58 W x 71 L cm (minimum guaranteed dimensions), minimum surface area 4,118 cm².
The side panels made of transparent acrylic resin are collapsible/foldable for complete accessibility.
Is equipped with a tiltable cradle, up to at least 10°.
Mattress covering the entire table surface, thickness no less than 5 cm.
Mattress cover flame-retardant, waterproof, and able to withstand hospital-grade disinfectant products.
Warm-up time < 30 min from 25°C.
OVERHEAD HEATER FEATURES:
Overhead radiant heater at a fixed height of 1.80m.
Radiant heating is achieved by quartz/ceramic elements, placed in a parabolic reflector.
Radiator is equipped with a safety grid.
The unit support three (3) modes: Pre-heating, manual and servo mode.
The unit allows for instant switching been the different operating modes.
The minimum skin temperature setting range is 32°C to 37°C.
The temperature control accuracy is: ± 0.5°C and the setting adjustment resolution is 0.1°C.
In manual and pre-heating modes heating settings allow for a range of 0 – 100%.
LED spotlight(s) for examination are integrated in the overhead radiant heater, the angle of the overhead light is adjustable.
| Pack Size | Avg. Price |
|---|---|